SM8004
Magnetic-inductive flow meter. SMR11GGX50KG/US100
Precise measurement of flow, consumption and medium temperature. High accuracy, repeatability and measurement dynamics. For use in water cooling circuits. With two precise analogue outputs. Clearly visible 4-digit LED display
Number of inputs and outputs: | Number of analogue outputs: 2 |
Process connection: | threaded connection G 1 DN25 flat seal |
Measuring range [°C]: | -20...80 |
Special feature: | Gold-plated contacts |
Application: | for industrial applications |
Installation: | connection to pipe by means of an adapter |
Media: | conductive liquids; water; hydrous media |
Note on media: | conductivity: ≥ 20 µS/cm viscosity: < 70 mm²/s (40 °C) |
Medium temperature [°C]: | -10...70 |
Pressure rating [bar]: | 16 |
Pressure rating [Mpa]: | 1.6 |
MAWP (for applications according to CRN) [bar]: | 11.2 |
Operating voltage [V]: | 20...30 DC; (according to EN 50178 SELV/PELV) |
Current consumption [mA]: | 120; (24 V) |
Protection class: | III |
Reverse polarity protection: | yes |
Power-on delay time [s]: | 5 |
Number of inputs and outputs: | Number of analogue outputs: 2 |
Total number of outputs: | 2 |
Output signal: | analogue signal |
Number of analogue outputs: | 2 |
Analogue current output [mA]: | 4...20; (scalable) |
Max. load [Ω]: | 500 |
Overload protection: | yes |
Measuring range: | 0.2...100 l/min 0.1...26.4 gpm |
0.2...100 l/min: | 0.1...26.4 gpm |
Display range: | -120...120 l/min -31.7...31.7 gpm |
-120...120 l/min: | -31.7...31.7 gpm |
Resolution: | 0.1 l/min 0.05 gpm |
0.1 l/min: | 0.05 gpm |
Analogue start point ASP: | 0...80 l/min 0...21.1 gpm |
0...80 l/min: | 0...21.1 gpm |
Analogue end point AEP: | 20...100 l/min 5.3...26.4 gpm |
20...100 l/min: | 5.3...26.4 gpm |
In steps of: | 0.1 l/min 0.05 gpm |
0.1 l/min: | 0.05 gpm |
Measuring range [°C]: | -20...80 |
Resolution [°C]: | 0.2 |
Analogue start point [°C]: | -20...60 |
Analogue end point [°C]: | 0...80 |
In steps of [°C]: | 0.2 |
Accuracy (in the measuring range): | ± (2 % MW + 0,5 % MEW) |
Repeatability: | ± 0,2% MEW |
Accuracy [K]: | ± 2,5 (Q > 1 l/min) |
Response time [s]: | 0.15; (dAP = 0, T19) |
Damping for the switching output dAP [s]: | 0...3 |
Dynamic response T05 / T09 [s]: | T09 = 20 (Q > 1 l/min) |
Ambient temperature [°C]: | -10...60 |
Storage temperature [°C]: | -25...80 |
Protection: | IP 67 |
EMC: | DIN EN 60947-5-9 500 V withstand voltage (V DC) |
DIN EN 60947-5-9: | 500 V withstand voltage (V DC) |
Shock resistance: | DIN IEC 68-2-27 20 g (11 ms) |
DIN IEC 68-2-27: | 20 g (11 ms) |
Vibration resistance: | DIN IEC 68-2-6 5 g (10...2000 Hz) |
DIN IEC 68-2-6: | 5 g (10...2000 Hz) |
MTTF [years]: | 175 |
Pressure Equipment Directive: | Sound engineering practice; can be used for group 2 fluids; group 1 fluids on request |
Weight [g]: | 618.5 |
Materials: | stainless steel (1.4404 / 316L); PBT-GF20; PC; FKM; TPE |
Materials (wetted parts): | stainless steel (1.4404 / 316L); PEEK; FKM |
Process connection: | threaded connection G 1 DN25 flat seal |
Display: | Display unit 6 x LED, green (l/min, m³/h, gpm, gph, °C, °F) measured values alphanumeric display, 4-digit programming alphanumeric display, 4-digit |
Display unit: | 6 x LED, green (l/min, m³/h, gpm, gph, °C, °F) |
measured values: | alphanumeric display, 4-digit |
programming: | alphanumeric display, 4-digit |
Display unit: | l/min; m³/h; gpm; gph; °C; °F |
Remarks: | MW = measured value MEW = Final value of the measuring range |
Pack quantity: | 1 pcs. |
Connection: | Connector: 1 x M12; Contacts: gold-plated |
ECCN: | |
Country of Origin: | DE |
EAN: | |
Weight: | 0.62 kg |
Length: | 0.00 m |
Width: | 0.00 m |
Height: | 0.00 m |