Number of inputs and outputs: | Number of digital outputs: 1; Number of analogue outputs: 1 |
Probe length L [mm]: | 100...1600 |
Process connection: | G 3/4 external thread |
Special feature: | Gold-plated contacts |
Media: | Liquids |
Dielectric constant of the medium: | ≥ 1,8; (for media with a dielectric constant of 1.8...5 (e.g. oils), a coaxial pipe is needed for operation) |
Recommended media: | water; hydrous media; oils; oil-based media |
Cannot be used for: | See the operating instructions, chapter "Function and features". |
Medium temperature [°C]: | -25...80; (90 < 1 h) |
Tank pressure [bar]: | -1...16 |
MAWP (for applications according to CRN) [bar]: | 16 |
Operating voltage [V]: | 18...30 DC |
Current consumption [mA]: | < 25 |
Protection class: | III |
Reverse polarity protection: | yes |
Power-on delay time [s]: | < 3 |
Number of inputs and outputs: | Number of digital outputs: 1; Number of analogue outputs: 1 |
Total number of outputs: | 2 |
Output signal: | analogue signal; IO-Link |
Electrical design: | PNP |
Number of digital outputs: | 1 |
Number of analogue outputs: | 1 |
Analogue current output [mA]: | 4...20, invertible |
Max. load [Ω]: | 500 |
Analogue voltage output [V]: | 0...10, invertible |
Min. load resistance [Ω]: | 2000 |
Short-circuit protection: | yes |
Type of short-circuit protection: | thermal, pulsed |
Overload protection: | yes |
Probe length L [mm]: | 100...1600 |
Active range A [mm]: | L-40; (when set to oil and oil based media: L-60) |
Inactive range I1 / I2 [mm]: | 30 / 10; (when set to oil and oil based media: 30 / 30) |
Sampling rate [Hz]: | 4 |
Repeatability [mm]: | ± 5 |
Measuring error [mm]: | ± 7 |
Offset error [mm]: | 5 |
Resolution [mm]: | 1 |
Zero signal (voltage) [V]: | 0 |
Zero signal (current) [mA]: | 4 |
Full signal (voltage) [V]: | 10 |
Full signal (current) [mA]: | 20 |
Temperature drift per 10 K: | ± 0.2 % |
Communication interface: | IO-Link |
Transmission type: | COM2 (38,4 kBaud) |
IO-Link revision: | 1.1 |
SDCI standard: | IEC 61131-9 CDV |
IO-Link device ID: | 578 d / 00 02 42 h |
Profiles: | no profile |
SIO mode: | no |
Required master port type: | A |
Process data analogue: | 1 |
Min. process cycle time [ms]: | 2.3 |
Ambient temperature [°C]: | -25...60 |
Storage temperature [°C]: | -40...85 |
Protection: | IP 68; IP 69K; (7 days / 1 m water depth / 0.1 bar: IP 68) |
EMC: | DIN EN 61000-6-2 DIN EN 61000-6-3 in a closed metal tank DIN EN 61000-6-4 in plastic or open metal tanks |
DIN EN 61000-6-3: | in a closed metal tank |
DIN EN 61000-6-4: | in plastic or open metal tanks |
Shock resistance: | DIN EN 60068-2-27 50 g (11 ms) / 25 g (6 ms) with reference rod 0.5 m |
DIN EN 60068-2-27: | 50 g (11 ms) / 25 g (6 ms) with reference rod 0.5 m |
Vibration resistance: | DIN EN 60068-2-6 5 g (10...2000 Hz) / 1 g (5...200 Hz) with reference rod 0.5 m |
DIN EN 60068-2-6: | 5 g (10...2000 Hz) / 1 g (5...200 Hz) with reference rod 0.5 m |
MTTF [years]: | 239 |
UL approval: | UL Approval no. H009 File number UL E174191 |
UL Approval no.: | H009 |
File number UL: | E174191 |
Weight [g]: | 430 |
Materials: | stainless steel (1.4301 / 304); stainless steel (1.4404 / 316L); FKM; PEI |
Materials (wetted parts): | stainless steel (1.4305 / 303); probe connection: stainless steel (1.4435 / 316L); PTFE; FKM; Sealing: NBR reinforced fibre |
Process connection: | G 3/4 external thread |
Notes: | Please see the technical note under "Downloads" |
Pack quantity: | 1 pcs. |
Connection: | Connector: 1 x M12; Contacts: gold-plated |
ECCN: | |
Country of Origin: | DE |
EAN: | |
Weight: | 0.43 kg |
Length: | 0.00 m |
Width: | 0.00 m |
Height: | 0.00 m |