TAD191
Temperature transmitter with drift detection. TAD088KLER12-A-DKG/US
| Number of inputs and outputs: | Number of digital outputs: 1; Number of analogue outputs: 1 |
| Measuring range: | -25...160 °C -13...320 °F |
| -25...160 °C: | -13...320 °F |
| Process connection: | threaded connection G 1/2 external thread |
| Installation length EL [mm]: | 87.5 |
| Special feature: | Gold-plated contacts |
| Measuring element: | 1 x Pt 1000 + 1 x NTC; (thermically coupled, with backup function (temperature measuring even if one element fails)) |
| Media: | liquids and gases |
| Pressure rating [bar]: | 50 |
| Minimum installation depth [mm]: | 25 |
| MAWP (for applications according to CRN) [bar]: | 50 |
| Operating voltage [V]: | 18...32 DC; ("supply class 2" to cULus) |
| Current consumption [mA]: | 6; (24 V) |
| Protection class: | III |
| Reverse polarity protection: | yes |
| Power-on delay time [s]: | 8 |
| Integrated watchdog: | yes |
| Number of inputs and outputs: | Number of digital outputs: 1; Number of analogue outputs: 1 |
| Total number of outputs: | 2 |
| Output signal: | switching signal; analogue signal; IO-Link; (configurable) |
| Electrical design: | PNP/NPN |
| Number of digital outputs: | 1 |
| Output function: | normally open / normally closed / diagnostic signal; (parameterisable) |
| Max. voltage drop switching output DC [V]: | 2 |
| Permanent current rating of switching output DC [mA]: | 150; (200 (...60°C); 250 (...40 °C)) |
| Diagnostic output: | diagnostic output for drift and fault monitoring |
| Number of analogue outputs: | 1 |
| Analogue current output [mA]: | 4...20 |
| Max. load [Ω]: | (Ub - 15 V) x 50 |
| Short-circuit protection: | yes |
| Type of short-circuit protection: | pulsed |
| Overload protection: | yes |
| Measuring range: | -25...160 °C -13...320 °F |
| -25...160 °C: | -13...320 °F |
| Note on measuring range: | scalable |
| Factory setting: | 0...150 °C / 32...302 °F |
| Drift warning: | 0.2...5 °C 0.4...9 °F |
| 0.2...5 °C: | 0.4...9 °F |
| Drift alarm: | 0.2...5 °C 0.4...9 °F |
| 0.2...5 °C: | 0.4...9 °F |
| In steps of: | 0.05 °C 0.1 °F |
| 0.05 °C: | 0.1 °F |
| Resolution of analogue output [K]: | 0.05 |
| Precision analogue output [K]: | ± 0,2 (-10...130°C); ± 0,3 (130...140°C); ± 0,3 ± 0,1% MS (-25...-10/140...160°C); (probe completely inserted into the measured medium up to the sealing chamfer) |
| Temperature coefficient [% of the span / 10 K]: | < ± 0,01; (in case of deviation from the reference condition 25 ± 5 °C) |
| Dynamic response T05 / T09 [s]: | 3 / 6 |
| Parameter setting options: | drift warning/drift alarm threshold; Fail-Safe; Display unit; scaling of the analogue output; redundancy switching; performance diagnostic output; switching logic; normally open / normally closed |
| Communication interface: | IO-Link |
| Transmission type: | COM1 (4,8 kBaud) |
| IO-Link revision: | 1.1 |
| SDCI standard: | IEC 61131-9 CDV |
| IO-Link device ID: | 323 d / 00 01 43 h |
| Profiles: | Smart Sensor: Process Data Variable; Device Identification, Device Diagnosis |
| SIO mode: | yes |
| Required master port type: | A |
| Process data analogue: | 1 |
| Process data binary: | 1 |
| Min. process cycle time [ms]: | 18.8 |
| Ambient temperature [°C]: | -25...70 |
| Storage temperature [°C]: | -40...85 |
| Protection: | IP 68; IP 69K |
| EMC: | DIN EN 61000-6-2 DIN EN 61000-6-3 |
| Shock resistance: | DIN EN 68000-2-27 50 g (11 ms) |
| DIN EN 68000-2-27: | 50 g (11 ms) |
| Vibration resistance: | DIN EN 60068-2-6 20 g (10...2000 Hz) |
| DIN EN 60068-2-6: | 20 g (10...2000 Hz) |
| MTTF [years]: | 213 |
| Note on approval: | Including free 5-point calibration certificate. |
| Weight [K40]: | 345 |
| Materials: | stainless steel (1.4435 / 316L); stainless steel (1.4404 / 316L); PEI; FKM |
| Materials (wetted parts): | stainless steel (1.4435 / 316L); stainless steel (1.4404 / 316L) |
| Tightening torque [Nm]: | 30...50 |
| Process connection: | threaded connection G 1/2 external thread |
| Surface characteristics Ra/Rz of the wetted parts: | Ra: < 0.6 |
| Installation length EL [mm]: | 87.5 |
| Remarks: | MS = set measuring span |
| Pack quantity: | 1 pcs. |
| Connection: | Connector: 1 x M12; Contacts: gold-plated |
| ECCN: | |
| Kraj pochodzenia: | US |
| EAN: | |
| Waga: | 0,35 kg |
| Długość: | 0,00 m |
| Szerokość: | 0,00 m |
| Wysokość: | 0,00 m |