Ifm SV5200

Vortex flow meter with display. SVR12XXXIRKG/US-100

Na prośbę. Prześlij nam wiadomość e-mail z zapytaniem o cenę i dostępność.
"ucnguBrctvdgcxgt0eqo"

Opis

Reliable detection of flow and medium temperature. With switching output or frequency output and IO-Link. Display can be rotated for optimum alignment. Clearly readable colour display with red/green colour change. Suitable for water, deionised water and cooling water

Parametry techniczne

Number of inputs and outputs: Number of digital outputs: 2
Process connection: threaded connection G 1/2 DN10
Measuring range [°C]: -10...90
Special feature: Gold-plated contacts
Application: for industrial applications
Media: water
Medium temperature [°C]: -10...90
Pressure rating [bar]: 12
Pressure rating [Mpa]: 1.2
Note on pressure rating: up to 40 °C
MAWP (for applications according to CRN) [bar]: 4.8
Operating voltage [V]: 18...30 DC
Current consumption [mA]: < 30
Min. insulation resistance [MΩ]: 100; (500 V DC)
Protection class: III
Reverse polarity protection: yes
Power-on delay time [s]: < 3
Number of inputs and outputs: Number of digital outputs: 2
Total number of outputs: 2
Output signal: switching signal; frequency signal; IO-Link; (configurable)
Electrical design: PNP/NPN
Number of digital outputs: 2
Output function: normally open / normally closed; (parameterisable)
Max. voltage drop switching output DC [V]: 2.5
Permanent current rating of switching output DC [mA]: 100
Short-circuit protection: yes
Overload protection: yes
Measuring range: 2...40 l/min 0.12...2.4 m³/h
2...40 l/min: 0.12...2.4 m³/h
Display range: 0...48 l/min 0...2.88 m³/h
0...48 l/min: 0...2.88 m³/h
Resolution: 0.2 l/min 0.01 m³/h
0.2 l/min: 0.01 m³/h
Set point SP: 2.4...40 l/min 0.14...2.4 m³/h
2.4...40 l/min: 0.14...2.4 m³/h
Reset point rP: 2...39.6 l/min 0.12...2.38 m³/h
2...39.6 l/min: 0.12...2.38 m³/h
Frequency end point, FEP: 8...40 l/min 0.48...2.4 m³/h
8...40 l/min: 0.48...2.4 m³/h
In steps of: 0.2 l/min 0.01 m³/h
0.2 l/min: 0.01 m³/h
Frequency at the end point FRP [Hz]: 100...1000
Measuring dynamics: 1:20
Measuring range [°C]: -10...90
Display range [°C]: -30...110
Resolution [°C]: 0.5
Set point SP [°C]: -9...90
Reset point rP [°C]: -10...89
In steps of [°C]: 0.5
Frequency start point, FSP [°C]: -10...70
Frequency end point, FEP [°C]: 10...90
Frequency at the end point FRP [Hz]: 100...1000
Accuracy (in the measuring range): ± 2 % MEW
Repeatability: ± 0,5 % MEW
Accuracy [K]: ± 1
Response time [s]: 1; (dAP = 0)
Damping for the switching output dAP [s]: 0...5
Dynamic response T05 / T09 [s]: T09 = 6
Parameter setting options: hysteresis / window; normally open / normally closed; switching logic; frequency output; switch-on/switch-off delay; Damping; Display unit
Communication interface: IO-Link
Transmission type: COM2 (38,4 kBaud)
IO-Link revision: 1.1
SDCI standard: IEC 61131-9
IO-Link device ID: 488 d / 00 01 E8 h
Profiles: Smart Sensor: Process Data Variable; Device Identification, Device Diagnosis
SIO mode: yes
Required master port type: A
Process data analogue: 2
Process data binary: 2
Min. process cycle time [ms]: 3
Ambient temperature [°C]: 0...60
Note on ambient temperature: medium temperature < 80 °C medium temperature < 90 °C: 0...50 °C
Storage temperature [°C]: -20...80
Protection: IP 65; IP 67
EMC: DIN EN 61000-6-2 DIN EN 61000-6-3
Shock resistance: DIN EN 60068-2-27 5 g (11 ms)
DIN EN 60068-2-27: 5 g (11 ms)
Vibration resistance: DIN EN 60068-2-6 with water / 10...50 Hz 1 mm with water / 50...2000 Hz 2 g
DIN EN 60068-2-6: with water / 10...50 Hz 1 mm
MTTF [years]: 342
UL approval: UL Approval no. I001
UL Approval no.: I001
Pressure Equipment Directive: Sound engineering practice; can be used for group 2 fluids; group 1 fluids on request
Weight [g]: 440
Materials: stainless steel (1.4404 / 316L); PC; PBT+PC-GF30; PPS; TPE-U
Materials (wetted parts): stainless steel (1.4404 / 316L); ETFE; PA 6T; PPS; FKM
Tightening torque [Nm]: 30
Process connection: threaded connection G 1/2 DN10
Remarks: MW = measured value MEW = Final value of the measuring range
Pack quantity: 1 pcs.
Connection: Connector: 1 x M12; Contacts: gold-plated

Parametry logistyczne

ECCN:
Kraj pochodzenia: US
EAN:
Waga: 0,44 kg
Długość: 0,00 m
Szerokość: 0,00 m
Wysokość: 0,00 m