Ifm SU8200

Ultrasonic flow meter. SUR11HGBFPKG/W/US-100-IPF

Na prośbę. Prześlij nam wiadomość e-mail z zapytaniem o cenę i dostępność.
"ucnguBrctvdgcxgt0eqo"

Opis

Reliable detection of flow and medium temperature. Suitable for water, oils and glycol solutions. For installation in pipes. With two programmable switching outputs. Clearly visible 4-digit LED display

Parametry techniczne

Number of inputs and outputs: Number of digital outputs: 2
Process connection: threaded connection G 1 flat seal
Measuring range [°C]: -10...80
Special feature: Gold-plated contacts
Application: for industrial applications
Installation: connection to pipe by means of an adapter
Media: water; glycol solutions; coolants; oil
Note on media: low-viscosity oils with viscosity: 7...40 mm²/s (40 °C) high-viscosity oils with viscosity: 30...68 mm²/s (40 °C)
Medium temperature [°C]: -10...80
Pressure rating [bar]: 16
Pressure rating [Mpa]: 1.6
Operating voltage [V]: 19...30 DC; (according to EN 50178 SELV/PELV)
Current consumption [mA]: 100
Min. insulation resistance [MΩ]: 100; (500 V DC)
Protection class: III
Reverse polarity protection: yes
Power-on delay time [s]: 10
Number of inputs and outputs: Number of digital outputs: 2
Total number of outputs: 2
Output signal: switching signal
Electrical design: PNP
Number of digital outputs: 2
Output function: normally open / normally closed; (parameterisable)
Max. voltage drop switching output DC [V]: 2
Permanent current rating of switching output DC [mA]: 250; (per output)
Short-circuit protection: yes
Type of short-circuit protection: pulsed
Overload protection: yes
Measuring range: 0...100 l/min 0...26.42 gpm
0...100 l/min: 0...26.42 gpm
Display range: 0...120 l/min 0...31.7 gpm
0...120 l/min: 0...31.7 gpm
Resolution: 0.1 l/min 0.02 gpm
0.1 l/min: 0.02 gpm
Set point SP: 0.2...100 l/min 0.06...26.42 gpm
0.2...100 l/min: 0.06...26.42 gpm
Reset point rP: 0...99.8 l/min 0...26.38 gpm
0...99.8 l/min: 0...26.38 gpm
Max. flow rate [l/min]: 110
In steps of: 0.1 l/min 0.02 gpm
0.1 l/min: 0.02 gpm
Measuring range [°C]: -10...80
Resolution [°C]: 0.2
Set point SP [°C]: -9.8...80
Reset point rP [°C]: -10...79.8
In steps of [°C]: 0.2
Accuracy (in the measuring range): < ± (3 % MW + 0,2 % MEW) / < ± (5 % MW + 0,5 % MEW); (water; glycol: 35%; oil: viscosity 68 mm²/s at 40 °C)
Repeatability: 0,2 l/min (0,05 gpm)
Accuracy [K]: 0,2
Response time [s]: 0.25; (dAP = 0)
Delay time programmable dS, dr [s]: 0...50
Damping for the switching output dAP [s]: 0...1
Dynamic response T05 / T09 [s]: T09 = 70 (Q > 5 l/min); (water)
Parameter setting options: Flow monitoring; Temperature monitoring
Ambient temperature [°C]: -10...60
Storage temperature [°C]: -25...80
Protection: IP 67
EMC: EN 61000-4-2 ESD 4 kV CD / 8 kV AD EN 61000-4-3 HF radiated 10 V/m EN 61000-4-4 Burst 2 kV EN 61000-4-5 Surge 0,5 kV EN 61000-4-6 HF conducted 10 V
EN 61000-4-2 ESD: 4 kV CD / 8 kV AD
EN 61000-4-3 HF radiated: 10 V/m
EN 61000-4-4 Burst: 2 kV
EN 61000-4-5 Surge: 0,5 kV
EN 61000-4-6 HF conducted: 10 V
Shock resistance: DIN IEC 68-2-27 20 g (11 ms)
DIN IEC 68-2-27: 20 g (11 ms)
Vibration resistance: DIN IEC 68-2-6 5 g (10...2000 Hz)
DIN IEC 68-2-6: 5 g (10...2000 Hz)
MTTF [years]: 246
Pressure Equipment Directive: Sound engineering practice; can be used for group 2 fluids; group 1 fluids on request
Weight [g]: 1694
Materials: housing: AlMgSi0,5 anodised; Sealing: FKM; connector housing: brass Optalloy-plated; PA 6.6; cover film: PA
Materials (wetted parts): stainless steel (1.4404 / 316L); FKM; PES; Centellen 200
Process connection: threaded connection G 1 flat seal
Display: Display unit 4 x LED, green (l/min, m³/h, gal/min, gal/h) Display unit 2 x LED, yellow (°C / °F) switching status 2 x LED, yellow measured values alphanumeric display, 4-digit programming alphanumeric display, 4-digit
Display unit: 4 x LED, green (l/min, m³/h, gal/min, gal/h)
Display unit: 2 x LED, yellow (°C / °F)
switching status: 2 x LED, yellow
measured values: alphanumeric display, 4-digit
programming: alphanumeric display, 4-digit
Display unit: l/min; m³/h; gpm; gph; °C; °F
Accessories (supplied): sealings: 2, Centellen
Accessories (optional): adapter for pipe: 1 x R 1/2, stainless steel, E40179 adapter for pipe: 1 x R 3/4, stainless steel, E40180 adapter for pipe: 1 x 1/2" NPT, stainless steel, E40192 adapter for pipe: 1 x 3/4" NPT, stainless steel, E40193 adapter for pipe: 1 x R 1/2, brass, E40152 adapter for pipe: 1 x R 3/4, brass, E40153 adapter for pipe: 1 x 1/2" NPT, brass, E40155 adapter for pipe: 1 x 3/4" NPT, brass, E40156
Remarks: MW = measured value MEW = Final value of the measuring range sealing: only with supplied Centellen seals
Pack quantity: 1 pcs.
Connection: Connector: 1 x M12; Moulded body: brass, Optalloy-plated; Contacts: gold-plated

Parametry logistyczne

ECCN:
Kraj pochodzenia: US
EAN:
Waga: 1,69 kg
Długość: 0,00 m
Szerokość: 0,00 m
Wysokość: 0,00 m