SD0523
Compressed air meter. SDD11DGXFPKG/US-100
Precise measurement of flow, consumption and medium temperature. High accuracy, repeatability and measurement dynamics. For detecting compressed air in industrial use. With switching output, analogue output and pulse output. Adjustable to inside pipe diameters
Number of inputs and outputs: | Number of digital outputs: 2; Number of analogue outputs: 1 |
Process connection: | threaded connection G 1 internal thread |
Measuring range [°C]: | 0...60 |
Installation: | Adjustable to inside pipe diameters; (38...254 mm) |
Media: | compressed air |
Medium temperature [°C]: | 0...60 |
Pressure rating [bar]: | 16 |
Pressure rating [Mpa]: | 1.6 |
Pressure rating [psi]: | 232 |
MAWP (for applications according to CRN) [bar]: | 16 |
Operating voltage [V]: | 18...30 DC; (according to EN 50178 SELV/PELV) |
Current consumption [mA]: | < 110 |
Protection class: | III |
Reverse polarity protection: | yes |
Power-on delay time [s]: | 1 |
Number of inputs and outputs: | Number of digital outputs: 2; Number of analogue outputs: 1 |
Total number of outputs: | 2 |
Output signal: | switching signal; analogue signal; pulse signal; IO-Link; (configurable) |
Electrical design: | PNP |
Number of digital outputs: | 2 |
Output function: | normally open / normally closed; (parameterisable) |
Max. voltage drop switching output DC [V]: | 2 |
Permanent current rating of switching output DC [mA]: | 250; (per output) |
Number of analogue outputs: | 1 |
Analogue current output [mA]: | 4...20; (scalable) |
Max. load [Ω]: | 500 |
Pulse output: | consumed quantity meter |
Short-circuit protection: | yes |
Type of short-circuit protection: | pulsed |
Overload protection: | yes |
Measuring range: | Measuring/setting range; Flow monitoring; The values apply at: Ø 72 mm |
Measuring range: | 0.5...143.9 m/s 8...2110 m³/h 0.12...35.18 m³/min |
0.5...143.9 m/s: | 8...2110 m³/h |
Display range: | 0...172.7 m/s 0...2532 m³/h 0...42.22 m³/min |
0...172.7 m/s: | 0...2532 m³/h |
Resolution: | 0.1 m/s 2 m³/h 0.02 m³/min |
0.1 m/s: | 2 m³/h |
Set point SP: | 1.2...143.9 m/s 18...2110 m³/h 0.28...35.18 m³/min |
1.2...143.9 m/s: | 18...2110 m³/h |
Reset point rP: | 0.5...143.2 m/s 6...2100 m³/h 0.12...35 m³/min |
0.5...143.2 m/s: | 6...2100 m³/h |
Analogue start point ASP: | 0...107.9 m/s 0...1582 m³/h 0...26.38 m³/min |
0...107.9 m/s: | 0...1582 m³/h |
Analogue end point AEP: | 36...143.9 m/s 528...2110 m³/h 8.8...35.18 m³/min |
36...143.9 m/s: | 528...2110 m³/h |
In steps of: | 0.1 m/s 2 m³/h 0.02 m³/min |
0.1 m/s: | 2 m³/h |
Pulse value: | 1...1000 x 10³ |
In steps of: | 1 Nm³ |
Pulse length [s]: | 0,7...2 (D = 72 mm) |
Measuring range [°C]: | 0...60 |
Display range [°C]: | -12...72 |
Resolution [°C]: | 0.2 |
Set point SP [°C]: | 0.4...60 |
Reset point rP [°C]: | 0.2...59.8 |
Analogue start point [°C]: | 0...45 |
Analogue end point [°C]: | 15...60 |
In steps of [°C]: | 0.2 |
Repeatability [% of the measured value]: | ± 1,5 |
Accuracy (in the measuring range): | ± (6 % MW + 0,6 % MEW); (D = 72mm, T = 22 °C; standard volume flow: 50...850 Nm³/h) |
Accuracy [K]: | ± 2,5 (Q > 2 Nm³/h) |
Response time [s]: | 0.1; (dAP = 0) |
Damping for the switching output dAP in steps [s]: | 0 - 0,2 - 0,4 - 0,6 - 0,8 - 1 |
Dynamic response T05 / T09 [s]: | 30 (Q > 2 Nm³/h) |
Parameter setting options: | Flow monitoring; quantity meter; Preset counter; Temperature monitoring; hysteresis / window; normally open / normally closed; current/pulse output; display can be rotated and switched off; Display unit; totaliser |
Communication interface: | IO-Link |
Transmission type: | COM2 (38,4 kBaud) |
IO-Link revision: | 1.1 |
SDCI standard: | IEC 61131-9 |
IO-Link device ID: | 381d / 00 01 7D h |
Profiles: | Smart Sensor: Process Data Variable; Device Identification |
SIO mode: | yes |
Required master port type: | A |
Process data analogue: | 3 |
Process data binary: | 2 |
Min. process cycle time [ms]: | 5 |
Ambient temperature [°C]: | 0...60 |
Storage temperature [°C]: | -20...85 |
Max. relative air humidity [%]: | 90 |
Protection: | IP 65 |
EMC: | DIN EN 60947-5-9 DIN EN 61000-6-3 |
Vibration resistance: | DIN EN 60068-2-6 5 g (55...2000 Hz) |
DIN EN 60068-2-6: | 5 g (55...2000 Hz) |
MTTF [years]: | 213 |
Weight [g]: | 442 |
Materials: | PBT-GF20; PC; stainless steel (1.4301 / 304); FKM |
Materials (wetted parts): | stainless steel (1.4401 / 316); stainless steel (1.4301 / 304); ceramics glass passivated; PEEK; polyester; FKM |
Process connection: | threaded connection G 1 internal thread |
Display: | Display unit 4 x LED, green (Nm³/min, Nm³/h, Nm³, °C) function display 1 x LED, green switching status 2 x LED, yellow measured values alphanumeric display, 4-digit programming alphanumeric display, 4-digit |
Display unit: | 4 x LED, green (Nm³/min, Nm³/h, Nm³, °C) |
function display: | 1 x LED, green |
switching status: | 2 x LED, yellow |
measured values: | alphanumeric display, 4-digit |
programming: | alphanumeric display, 4-digit |
Remarks: | Measuring, display and setting ranges refer to the standard volume flow according to DIN ISO 2533. D = inside pipe diameter |
Pack quantity: | 1 pcs. |
Connection: | Connector: 1 x M12 |
ECCN: | |
Kraj pochodzenia: | DE |
EAN: | |
Waga: | 0,44 kg |
Długość: | 0,00 m |
Szerokość: | 0,00 m |
Wysokość: | 0,00 m |