Measurement of flow and temperature at the same time. Optimised for water, oils and air. Clearly visible 4-digit LED display. With two scalable analogue outputs. The process connection can be rotated for optimum alignment

Parametry techniczne

Number of inputs and outputs: Number of analogue outputs: 2
Process connection: threaded connection M18 x 1,5 internal thread
Special feature: Gold-plated contacts
Media: water; glycol solutions; air; oils
Note on media: low-viscosity oils with viscosity: ≤ 40 mm²/s (40 °C) high-viscosity oils with viscosity: > 40 mm²/s (40 °C)
Medium temperature [°C]: -20...90
Pressure rating [bar]: 100
MAWP (for applications according to CRN) [bar]: 100
Operating voltage [V]: 18...30 DC
Current consumption [mA]: < 100
Protection class: III
Reverse polarity protection: yes
Power-on delay time [s]: 10
Number of inputs and outputs: Number of analogue outputs: 2
Total number of outputs: 2
Output signal: analogue signal
Number of analogue outputs: 2
Analogue current output [mA]: 4...20; (scalable)
Max. load [Ω]: 350
Short-circuit protection: yes
Type of short-circuit protection: pulsed
Overload protection: yes
Probe length L [mm]: 45
Operating mode: relative; absolutely liquid; absolutely gaseous; (absolute: reference measurement recommended)
Note on factory setting: Operating mode: relative
Setting range [m/s]: 0.04...6
Greatest sensitivity [m/s]: 0.04...3
Setting range [m/s]: 0...200
Greatest sensitivity [m/s]: 2...100
Measuring range [°C]: -20...90
Resolution [°C]: 0.2
Analogue start point [°C]: -20...76
Analogue end point [°C]: 4...100
In steps of [°C]: 0.2
Temperature drift [cm/s x 1/K]: 0,003 m/s x 1/K (< 20 °C; > 70 °C)
Temperature gradient [K/min]: 100
Accuracy: ± (7 % MW + 2 % MEW); (for relative mode in the range of maximum sensitivity under the following conditions :; water: 20...70 °C; inlet length: 1.5 m; DN25 (DIN 2448); mounting position according to instructions; Accuracy can differ for other media and mounting positions.)
Repeatability: 0,05 m/s; (water; flow velocity: 0,05...3 m/s)
Temperature drift: ± 0,005 K/°C
Accuracy [K]: ± 0,3 / ± 1; (water; flow velocity: 0,3...3 m/s / air; flow velocity: > 10 m/s)
Response time [s]: 0.5; (T09; water; glycol: 0,8 s; air: 7 s; oil: 1,8 s; each T09)
Dynamic response T05 / T09 [s]: 1,5 (T09); (water; flow velocity: 0,3...3 m/s)
Parameter setting options: medium selection; Damping; Teach function; display can be rotated and switched off; standard unit of measurement; process value colour
Ambient temperature [°C]: -40...80
Storage temperature [°C]: -40...100
Protection: IP 65; IP 67
EMC: DIN EN 60947-5-9
Shock resistance: DIN EN 60068-2-27 50 g (11 ms)
DIN EN 60068-2-27: 50 g (11 ms)
Vibration resistance: DIN EN 60068-2-6 20 g (10...2000 Hz)
DIN EN 60068-2-6: 20 g (10...2000 Hz)
MTTF [years]: 180
UL approval: UL Approval no. I004 File number UL E174189
UL Approval no.: I004
File number UL: E174189
Weight [g]: 276
Materials: stainless steel (1.4404 / 316L); stainless steel (1.4310 / 301); PBT-GF20; PBT-GF30
Materials (wetted parts): stainless steel (1.4404 / 316L); Gasket: FKM
Process connection: threaded connection M18 x 1,5 internal thread
Display: Display unit 6 x LED, green (%, m/s, l/min, m³/h, °C, 10³) measured values alphanumeric display, red/green 4-digit
Display unit: 6 x LED, green (%, m/s, l/min, m³/h, °C, 10³)
measured values: alphanumeric display, red/green 4-digit
Remarks: MW = measured value MEW = Final value of the measuring range
Pack quantity: 1 pcs.
Connection: Connector: 1 x M12; Contacts: gold-plated

Parametry logistyczne

ECCN:
Kraj pochodzenia: DE
EAN:
Waga: 0,28 kg